Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Simpson 260-3: CAL ADJ /5520 DATE: 09-Sep-97 AUTHOR: Fluke Corporation REVISION: $Revision: 1.1 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 1 NUMBER OF LINES: 97 CONFIGURATION: Fluke 5520A ============================================================================= # # Compatibility: # 5500/CAL 5.0 or MET/CAL 5.0 or later. # # Subprocedures: # None # # Required Files: # sim260a1.bmp # sim260a2.bmp # sim260a3.bmp # sim260a4.bmp # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5520A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- V 1.002 HEAD PRELIMINARY INSTRUCTIONS 1.003 DISP Warm-up time: None 1.003 DISP Temperature: 20 - 40degC 1.003 DISP Humidity: <80% 1.004 DISP [32] WARNING 1.004 DISP HIGH VOLTAGE is used or exposed during the performance 1.004 DISP of this calibration. DEATH ON CONTACT may result if 1.004 DISP personnel fail to observe safety precautions. 1.005 DISP [32] NOTICE 1.005 DISP Check the meter sensitivity before replacing components 1.005 DISP or performing calibration. A change in meter sensi- 1.005 DISP tivity is evident when all ranges are off calibration. 1.006 HEAD DISASSEMBLY INSTRUCTIONS 1.007 DISP Turn the FUNCTION SWITCH to +DC, and remove all 1.007 DISP connections and test leads. 1.007 DISP 1.007 DISP Remove the four screws securing the rear cover to the 1.007 DISP the front panel. 1.007 DISP 1.007 DISP Remove the rear cover. 1.008 HEAD DC CURRENT 1.009 DISP Adjust the front panel mechanical METER ZERO with the 1.009 DISP UUT in a horizontal position. 1.009 DISP 1.009 DISP Check that the adjuster pin does NOT rest against the 1.009 DISP adjustment fork when finished. 1.010 DISP Set the Function Switch to +DC. 1.010 DISP Set the Range Switch to 50uA. 1.011 DISP Connect the 5520A to the UUT as follows: 1.011 DISP [32] 5520A AUX HI to UUT 50uA 1.011 DISP [32] 5520A AUX LO to UUT COMMON 1.012 5520 50.0uA S 2W 1.013 PIC SIM260a1 1.014 5520 * S 1.015 HEAD DC VOLTAGE 1.016 DISP [32] * NOTE * 1.016 DISP It may be necessary to make a compromise adjustment 1.016 DISP of R27 to meet tolerances of both voltage and current 1.016 DISP ranges. Accuracy of the DC Voltage depends on the 1.016 DISP accuracy of the previous adjustment. 1.017 DISP Connect the 5520A to the UUT as follows: 1.017 DISP [32] 5520A NORMAL HI to UUT 50uA 1.017 DISP [32] 5520A NORMAL LO to UUT COMMON 1.018 5520 0.250V S 2W 1.019 PIC SIM260a2 1.020 5520 * S 1.021 DISP [32] *NOTE* 1.021 DISP A compromise will have to be reached between the 50uA 1.021 DISP & 250mV adjustments. Adjust for best overall accuracy. 1.022 OPBR Do you wish to re-do the 50uA adjustment? 1.023 JMPT 1.010 1.024 HEAD AC VOLTAGE 1.025 DISP Connect the 5520A to the UUT as follows: 1.025 DISP [32] 5520A NORMAL HI to UUT POS 1.025 DISP [32] 5520A NORMAL LO to UUT COMMON 1.026 DISP Set the Function Switch to AC. 1.026 DISP Set the Range Switch to 250V. 1.027 5520 250V 60H SI S 2W 1.028 PIC SIM260a3 1.029 5520 * S 1.030 DISP Set the Range Switch to 2.5V. 1.031 5520 2.50V 60H SI S 2W 1.032 PIC SIM260a4 1.033 END